CV Mode Calibration (Calibration numbers 4 and 5)
7
M
a
in
tenance
■
Step E: Calibrate the offset values of the internal reference voltage for
voltage settings and the measured values.
1
Use
and
to select “4. CV 65V.”
2
Connect a CC power supply to the load input terminals, and supply 0.05
A. Set the power supply voltage to 66 V or greater.
3
Press ENTER.
The load turns on automatically, and the offset calibration (CV 65V Offset Adjustment)
screen appears.
4
Press or to set the cursor position of DAC REF, and then turn the
rotary knob so that the input voltage is within ±0.05 % of the value
corresponding to 10 % of the full scale value (see the table of CV mode
settings).
The measured value offset MON is set automatically.
■
Step F: Calibrate the offset value for the internal reference voltage for the
protection functions.
5
Press ENTER.
The offset calibration screen (CV 65 V Limit Offs Adjust) appears.
6
Press or to set the cursor position of DAC LIM, and then turn the
rotary knob so that the input voltage is within ±0.05 % of the value
corresponding to 10 % of the full scale value (see the table of CV mode
settings).
7
Press ENTER.
The load turns off automatically.
■
Step G: Calibrate the gain values of the internal reference voltage for
voltage settings and of the measured values.
8
Press ENTER.
The load turns on automatically, and the gain calibration (CV 65V Gain Adjustment)
screen appears.
9
Press or to set the cursor position of DAC REF, and then turn the
rotary knob so that the input voltage is within ±0.05 % of the value
corresponding to 100 % of the full scale value (see the table of CV mode
settings).
The measured value gain MON is set automatically.
■
Step H: Calibrate the gain value for the internal reference voltage for the
protection function settings.
10
Press ENTER.
The gain calibration screen (CV 65 V Limit Gain Adjust) appears.
Offset Adjustment
PREV
NEXT
DAC REF : 0x0001
MON
:
0x0001
CV 4V
Cursor
Indicates that the values are hexadecimal
The current calibration value (hexadecimal)
Содержание PLZ1004WH
Страница 13: ...Installation and Preparation This chapter describes how to unpack and prepare this product before you use it 1 ...
Страница 28: ...28 PLZ 4WH Remote Sensing ...
Страница 29: ...Basic Functions This chapter explains the operations for each operation mode and other basic functions 2 ...
Страница 47: ...Advanced Operations This chapter explains ABC preset memory setup memory protection functions and the menu 3 ...
Страница 61: ...Sequences This chapter explains the sequence function Sequences 4 ...
Страница 83: ...External Control and Parallel Operation This chapter explains external control and Parallel Operation 5 ...
Страница 115: ...Maintenance This chapter describes the maintenance procedures including cleaning inspection and calibration 7 ...
Страница 127: ...8 Specifications This chapter contains the specifications and gives the dimensions of the PLZ 4WH ...
Страница 137: ...Appendix A Operating Area B Sequence Program Creation Table C Options ...