132
PLZ-4WH
Specifications
Soft start
Response speed
Remote sensing
Protection function
Model
PLZ164WH
PLZ334WH
PLZ1004WH
Operation mode
CC
Selectable times
1
1
The time it takes to shift from 10 % to 90 % of the rated current
1 ms, 2 ms, 5 ms, 10 ms, 20 ms, 50 ms, 100 ms, 200 ms
Time accuracy
±(30 % of set +100 µs)
Model
PLZ164WH
PLZ334WH
PLZ1004WH
CC mode, CR mode
1/1, 1/2, 1/5, 1/10
CV mode
100, 10, 1, 1/10, 1/100
Model
PLZ164WH
PLZ334WH
PLZ1004WH
Voltage that can be compensated
2 V for a single line
Model
PLZ164WH
PLZ334WH
PLZ1004WH
Overvoltage protection
(OVP)
Turns off the load when 110 % of the range’s maximum voltage is
detected.
Overcurrent protection
(OCP)
Can be set to 110 % of each range’s maximum current or the
values given below.
0.01 A to 9.07 A
0.01 A to 18.15 A
0.01 A to 55 A
The action can be set to load off or limit.
Overpower protection
(OPP)
Can be set to 110 % of each range’s maximum power or 0.1 % to
110 % of the rated power.
The action can be set to load off or limit.
Overheat protection (OHP) Turns off the load when the heat sink temperature reaches 90
°
C
Undervoltage protection
(UVP)
Turns off the load when the specified value is detected
Can be set to a value from 5 V to 650 V of OFF
Reverse-connection
protection (REV)
Implemented through a fuse and a diode
Turns off the load when an alarm occurs
Содержание PLZ1004WH
Страница 13: ...Installation and Preparation This chapter describes how to unpack and prepare this product before you use it 1 ...
Страница 28: ...28 PLZ 4WH Remote Sensing ...
Страница 29: ...Basic Functions This chapter explains the operations for each operation mode and other basic functions 2 ...
Страница 47: ...Advanced Operations This chapter explains ABC preset memory setup memory protection functions and the menu 3 ...
Страница 61: ...Sequences This chapter explains the sequence function Sequences 4 ...
Страница 83: ...External Control and Parallel Operation This chapter explains external control and Parallel Operation 5 ...
Страница 115: ...Maintenance This chapter describes the maintenance procedures including cleaning inspection and calibration 7 ...
Страница 127: ...8 Specifications This chapter contains the specifications and gives the dimensions of the PLZ 4WH ...
Страница 137: ...Appendix A Operating Area B Sequence Program Creation Table C Options ...