PLZ-4WH
121
CC Mode Calibration (Calibration numbers 1, 2, and 3)
7
M
a
in
tenance
5
Connect a constant-voltage power supply to the load input terminals,
and apply 5 V. Set the power supply current to approxi2 % to 5
% of the load device’s rated current.
6
Press ENTER.
The load turns on automatically, and the offset calibration (CC (Low) Offset
Adjustment) screen appears.
7
Press or to set the cursor position of DAC REF, and then turn the
rotary knob so that the current flowing through the shunt resistor is
within ±0.1 % of the value corresponding to 10 % of the full scale value
(see the table of CC mode settings).
The measured value offset MON is set automatically.
■
Step B: Calibrate the offset value for the internal reference voltage for the
protection functions.
8
Press ENTER.
The offset calibration screen (CC (Low) Limit Offs Adjust) appears.
9
Press or to set the cursor position of DAC LIM, and then turn the
rotary knob so that the current flowing through the shunt resistor is
within ±0.1 % of the value corresponding to 10 % of the full scale value
(see the table of CC mode settings).
10
Press ENTER.
The load turns off automatically.
■
Step C: Calibrate the gain values of the internal reference voltage for
current settings and of the measured values.
11
Connect a shunt resistor that matches the value corresponding to 100 %
of the maximum low range value.
12
Press ENTER.
The load turns on automatically, and the gain calibration screen (CC (Low) Gain
Adjustment) appears.
13
Press or to set the cursor position of DAC REF, and then turn the
rotary knob so that the current flowing through the shunt resistor is
within ±0.1 % of the value corresponding to 100 % of the full scale value
(see the table of CC mode settings).
The measured value gain MON is set automatically.
■
Step D: Calibrate the gain value for the internal reference voltage for the
protection function settings.
14
Press ENTER.
The gain calibration screen (CC (Low) Limit Gain Adjust) appears.
15
Press or to set the cursor position of DAC LIM, and then turn the
rotary knob so that the current flowing through the shunt resistor is
Offset Adjustment
PREV
NEXT
DAC REF : 0x0001
MON
:
0x0001
CC(Low)
Cursor
Indicates that the values are hexadecimal
The current calibration value(hexadecimal)
Содержание PLZ1004WH
Страница 13: ...Installation and Preparation This chapter describes how to unpack and prepare this product before you use it 1 ...
Страница 28: ...28 PLZ 4WH Remote Sensing ...
Страница 29: ...Basic Functions This chapter explains the operations for each operation mode and other basic functions 2 ...
Страница 47: ...Advanced Operations This chapter explains ABC preset memory setup memory protection functions and the menu 3 ...
Страница 61: ...Sequences This chapter explains the sequence function Sequences 4 ...
Страница 83: ...External Control and Parallel Operation This chapter explains external control and Parallel Operation 5 ...
Страница 115: ...Maintenance This chapter describes the maintenance procedures including cleaning inspection and calibration 7 ...
Страница 127: ...8 Specifications This chapter contains the specifications and gives the dimensions of the PLZ 4WH ...
Страница 137: ...Appendix A Operating Area B Sequence Program Creation Table C Options ...