24
PLZ-4WH
Connecting to the Load Input Terminals
Connection procedure
1
Turn the POWER switch off.
2
Make sure that the output of the DUT is off.
3
Attach the lock plates to the load input terminals on the rear panel.
After you attach the lock plates once, there is no need to remove them.
4
Attach crimping terminals to the load wiring.
The DC INPUT terminals on the rear panel have open bolt holes (M8) for connecting
the load wiring. Attach appropriate crimping terminals.
5
Pass the wiring that you want to connect to the load through the
attached load input terminal cover.
6
Connect the load wires to the load input terminals on the rear panel
using the included load input terminal screw set.
Lock plate
Insert the lock plates into
the slots by the terminals.
Bolt (M8 × 18)
Spring washer (M8)
Load input terminal cover
Nut (M8)
Crimping terminal
Be sure to use the
included screw set.
Содержание PLZ1004WH
Страница 13: ...Installation and Preparation This chapter describes how to unpack and prepare this product before you use it 1 ...
Страница 28: ...28 PLZ 4WH Remote Sensing ...
Страница 29: ...Basic Functions This chapter explains the operations for each operation mode and other basic functions 2 ...
Страница 47: ...Advanced Operations This chapter explains ABC preset memory setup memory protection functions and the menu 3 ...
Страница 61: ...Sequences This chapter explains the sequence function Sequences 4 ...
Страница 83: ...External Control and Parallel Operation This chapter explains external control and Parallel Operation 5 ...
Страница 115: ...Maintenance This chapter describes the maintenance procedures including cleaning inspection and calibration 7 ...
Страница 127: ...8 Specifications This chapter contains the specifications and gives the dimensions of the PLZ 4WH ...
Страница 137: ...Appendix A Operating Area B Sequence Program Creation Table C Options ...