40
PLZ-4WH
Switching
The switching function is used to switch between two preset load currents. Switching can be
performed in CC mode and CR mode.
Switching is suitable for transient-response testing of regulated DC power supplies and
similar devices.
To use the switching function, you need to set a switching level and a switching interval and
then turn on the switching function. You can configure switching while the load is on or off.
When you execute switching, a trigger signal is produced from the TRIG OUT terminal on the
front panel.
Example in CC Mode
Setting the switching level
You can set the switching level to a current or conductance value or to a percentage.
■
Setting the switching level to a current or conductance value
1
Press LEVEL.
The LEVEL key lights, and in CC mode, the current can be set; in CR mode, the
conductance can be set.
2
Turn the rotary knob to set the switching level.
■
Setting the switching level to a percentage
The specified current or conductance level is equivalent to 100 %.
1
Press % (SHIFT+LEVEL).
The LEVEL key lights, and you can set the switching level to a percentage.
2
Turn the rotary knob to set the switching level (0.0 % to 100.0 %).
Th
TL
FREQ
0[A]
(0 %)
LEVEL[A][S] [%]
SET[A]
(100 %)
Edge at which the pulse is output from the TRIG OUT connector
LEVEL: Switching level
FREQ: Frequency of the switching
interval
Th:
HIGH time of the switching
interval.
TL:
LOW time of the switching
interval.
SWITCHING
ENTER
SHORT
EDIT
SLEW RATE
RUN / STOP
UVP
PAUSE
STORE
RECALL
A
B
C
LOAD
Th/TL
FREQ / DUTY
MENU
PRESET SEQ
I
LEVEL
VRANGE
RANGE
SW ON
+CV
MODE
LOAD
SET / VSET
LOCK
SHIFT
ABC
OPP / OCP
Содержание PLZ1004WH
Страница 13: ...Installation and Preparation This chapter describes how to unpack and prepare this product before you use it 1 ...
Страница 28: ...28 PLZ 4WH Remote Sensing ...
Страница 29: ...Basic Functions This chapter explains the operations for each operation mode and other basic functions 2 ...
Страница 47: ...Advanced Operations This chapter explains ABC preset memory setup memory protection functions and the menu 3 ...
Страница 61: ...Sequences This chapter explains the sequence function Sequences 4 ...
Страница 83: ...External Control and Parallel Operation This chapter explains external control and Parallel Operation 5 ...
Страница 115: ...Maintenance This chapter describes the maintenance procedures including cleaning inspection and calibration 7 ...
Страница 127: ...8 Specifications This chapter contains the specifications and gives the dimensions of the PLZ 4WH ...
Страница 137: ...Appendix A Operating Area B Sequence Program Creation Table C Options ...