22
PLZ-4WH
Methods for Ensuring Stable Operation
Do not apply excessive voltages to the load input terminals
The maximum voltage that can be applied to the load input terminals is 650 Vdc. You cannot
use voltages that exceed 650 Vdc. When excessive voltage is applied, the protection
functions are activated. Lower the voltage of the DUT immediately.
Match the polarities of the load input terminals and the DUT
terminals
Connect each load input terminal to the terminal on the DUT with the same polarity.
The protection functions are activated when a reverse voltage of approximately 0.4 V or
greater is applied. If this happens, turn off the DUT immediately.
CAUTION
Do not apply a voltage that is greater than the maximum voltage of 650 Vdc to the load input
terminals. Doing so may damage the product.
100.00
A
150.00
V
16.50
W
CC
0.000
A
SET
SLEWRATE
2.50 m
A /µs
ALARM
ENTER
OVP
H 8.25A 650V
CAUTION
Connecting the electronic load to the DUT with the polarities reversed can result in the flow
of excessive current and damage to the DUT and the electronic load.
Device
under test
PLZ-4WH Series
Electronic Load
≤
650 Vdc
+
+
0.00
mA
0.000
V
0.000
W
CC
0.000
A
SET
ALARM
ENTER
SLEWRATE
25.0
µA /µs
Reverse
H 8.25A 650V
Содержание PLZ1004WH
Страница 13: ...Installation and Preparation This chapter describes how to unpack and prepare this product before you use it 1 ...
Страница 28: ...28 PLZ 4WH Remote Sensing ...
Страница 29: ...Basic Functions This chapter explains the operations for each operation mode and other basic functions 2 ...
Страница 47: ...Advanced Operations This chapter explains ABC preset memory setup memory protection functions and the menu 3 ...
Страница 61: ...Sequences This chapter explains the sequence function Sequences 4 ...
Страница 83: ...External Control and Parallel Operation This chapter explains external control and Parallel Operation 5 ...
Страница 115: ...Maintenance This chapter describes the maintenance procedures including cleaning inspection and calibration 7 ...
Страница 127: ...8 Specifications This chapter contains the specifications and gives the dimensions of the PLZ 4WH ...
Страница 137: ...Appendix A Operating Area B Sequence Program Creation Table C Options ...