Turning the Power On
Turning the POWER switch on
If the POWER switch is turned on for the first time after purchasing the PLZ-4WH, the PLZ-
4WH starts up using factory default settings. For all other cases, the PLZ-4WH starts up using
the settings that existed when the POWER switch was turned off the last time.
1
Check that the POWER switch is turned off ( ).
2
Check that the power cord is correctly connected.
3
Check that nothing is connected to the DC INPUT (load input terminal)
on the front panel and rear panel.
4
Turn the POWER switch on ( ).
5
Check that the display is in the fundamental setting entry condition.
The measured value displayed (section with mA, V, and W unit) indicates coarse zero.
The characters “SET” shown under the measured value is highlighted with an
underline. You can enter the fundamental setting for the selected operation mode.
6
Push LOAD.
Check that the LED above the key illuminates.
7
Push LOAD again.
Check that the LED above the key turns off.
If an odd sound, odd odor, fire, or smoke occurs around or in the PLZ-4WH, remove the
power plug from the outlet or turn off the power switch.
POWER switch
LOAD key
Load-on indicator LED
Memo
The condition in which
characters “SET” is
highlighted is called the
fundamental setting entry
condition.
Constant current (CC) mode is selected
in this screen, so you can enter the current,
which is the fundamental setting.
0.00
mA
0.000
V
0.000
W
CC
SLEWRATE
3.0000
A
0.01
mA/µs
SET
H 8.25A 650V
Содержание PLZ1004WH
Страница 13: ...Installation and Preparation This chapter describes how to unpack and prepare this product before you use it 1 ...
Страница 28: ...28 PLZ 4WH Remote Sensing ...
Страница 29: ...Basic Functions This chapter explains the operations for each operation mode and other basic functions 2 ...
Страница 47: ...Advanced Operations This chapter explains ABC preset memory setup memory protection functions and the menu 3 ...
Страница 61: ...Sequences This chapter explains the sequence function Sequences 4 ...
Страница 83: ...External Control and Parallel Operation This chapter explains external control and Parallel Operation 5 ...
Страница 115: ...Maintenance This chapter describes the maintenance procedures including cleaning inspection and calibration 7 ...
Страница 127: ...8 Specifications This chapter contains the specifications and gives the dimensions of the PLZ 4WH ...
Страница 137: ...Appendix A Operating Area B Sequence Program Creation Table C Options ...