Series 2600 System SourceMeters Reference Manual
Status Model D-29
Return to
2600S-901-01 Rev. A / May 2006
Measurement Event Reading Overflow Summary
This summary register set feeds to ROF bit B7 of the Measurement Event
Register. Used summary bits for this register include B1 (SMUA) and B2 (SMUB).
A set bit indicates that a reading overflow has occurred for a SMU (A or B).
Measurement Event Buffer Available Summary
This summary register set feeds to BAV bit B8 of the Measurement Event
Register. Used summary bits for this register include B1 (SMUA) and B2 (SMUB).
A set bit indicates that a buffer is available for a SMU (A or B).
Measurement Event SMU A Summary Register
This SMU summary register set feeds to bit B1 (SMUA) of the Measurement
Status Instrument Summary Register. Used summary bits for this register include
B0 (VLMT), B1 (ILMT), B7 (ROF) and B8 (BAV). A set bit indicates that the
corresponding measurement event for SMU A has occurred.
Measurement Event SMU B Summary Register
This SMU summary register set feeds to bit B2 (SMUB) of the Measurement
Status Instrument Summary Register. Used summary bits for this register include
B0 (VLMT), B1 (ILMT), B7 (ROF) and B8 (BAV). A set bit indicates that the
corresponding measurement event for SMU B has occurred.
Measurement Event Instrument Summary Register
This summary register is fed from the SMU summary registers and then feeds to
bit B13 (INST) of the Measurement Event Summary Register. Used summary bits
for this register include B1 (SMUA) and B2 (SMUB). A set bit indicates that an
measurement event has occurred for SMU A and/or SMU B.