Series 2600 System SourceMeters Reference Manual
Source-Measure Concepts 8-33
Return to
2600S-901-01 Rev. A / May 2006
Figure 8-17
Pulse rise and fall times
0V or 0A
90%
10%
Rise
Time
Fall
Time
Times Exaggerated
For Clarity
Programmed Fixed
or Sweep Step Level
Pulse duty cycle
Duty cycle is the percentage of time during the pulse period that the output is on.
It is calculated as follows:
Duty cycle = Pulse width / (Pulse width + Off-time)
For example, if the pulse width is 10msec, and the off-time is 90msec, the duty
cycle is calculated as follows:
Duty cycle= 10msec / ( 90msec)
= 10msec / 100msec
= 0.10
= 10%