Series 2600 System SourceMeters Reference Manual
Sweep Operation 5-11
Return to
2600S-901-01 Rev. A / May 2006
Running sweeps
Front panel
To run a sweep, press the LOAD key, then select the test to run. Press the RUN
key, then follow the display prompts to complete the test (refer to
through
for more information on
using factory scripts.
Press the RECALL key to access sweep data stored in Buffer 1. See
for
more details on the buffer.
Sweep programming examples
Procedures for programming and running a sweep for three sweep types are given
on the following pages. Each of these procedures includes commands for a typical
sweep example.
summarizes parameters for each of these examples.
Table 5-5
Sweep example parameters
Sweep type
Parameters for sweep examples
Linear staircase sweep
Start current: 1mA
Stop current: 10mA
# points: 10
Settling time: 0.1s
Pulse current sweep
Bias current: 1mA
On current: 10mA
Pulse on time: 10ms
Pulse off time: 50ms
# points: 10
Custom (list) sweep
Five points: 3V, 1V, 4V, 5V, 2V
Settling time 0.1s