Series 2600 System SourceMeters Reference Manual
Source-Measure Concepts 8-11
Return to
2600S-901-01 Rev. A / May 2006
Figure 8-3
Model 2601/2602 continuous power operating boundaries
-V
+V
-40V
(IV)
Sink
+I
3A
1A
-6V
6V
40V
(III)
Source
(II)
Sink
-1A
-3A
-I
(I)
Source
-2.2A
-600mA
600mA
2.2A
Energy storage DUT
Energy storage DUT
£
30°C
T
amb
Model 2611/2612 continuous power operating boundaries
The general operating boundaries for Model 2611/2612 continuous power output
are shown in
(see “
General SourceMeter power equation
” in this sec-
tion for derating factors). In this drawing, the 1.5A, 20V and 100mA, 200V
magnitudes are nominal values. Also note that the boundaries are not drawn to
scale.
Содержание SourceMeter 2600 Series
Страница 507: ...AppendixA Specifications Appendix A topics Model 2600S Specifications page A 2 2600S 901 01 Rev A May 2006...
Страница 575: ...AppendixF DisplayCharacterCodes Appendix F topics Introduction page F 2...
Страница 592: ......
Страница 593: ......