Series 2600 System SourceMeters Reference Manual
DUT Test Connections 3-17
Return to
2600S-901-01 Rev. A / May 2006
Figure 3-16
V-source
HI
LO
Series 2600
SMU
10V
+10V
Chassis Earth Ground
Low
Output low connected
to chassis
Series 2600 LO
NOT connected
to chassis ground
(floating)
High
+
Chassis
Chassis
Source chassis connected to chassis
earth ground through the power cord
External Source
Series 2600 chassis connected
to chassis earth ground
through the power cord
+
Floating the Series 2600
Figure 3-17
DUT
Low
10V
+
Chassis
External Source
Output low connected
to chassis
Source chassis connected to chassis
earth ground through the power cord
High
Keithley Model 2602/2612
Channel A
LO G
LO
HI G G G
HI
S
S
Channel B
LO
G
LO
HI
G
G
G
HI
S
S
HI
LO
Chassis
Screw
Low-Noise
Chassis Ground
Banana Jack
SMU connections for the floating configuration shown in