Table 15-8
Model 2601/2602 current measurement accuracy limits
Model 2601/2602
source and
measure range
1
Source
current
2
Model 2601/2602
current reading limits
(1 year, 18°C–28°C)
100nA
1µA
10µA
100µA
1mA
10mA
100mA
1A
3A
90.000nA
0.9000µA
9.0000µA
90.000µA
0.9000mA
9.0000mA
90.000mA
0.90000A
2.4000A
89.855 to 90.145nA
0.899475 to 0.900525µA
8.99625 to 9.00375µA
89.957 to 90.043µA
0.89962 to 0.90038mA
8.9957 to 9.0043mA
89.962 to 90.038mA
0.89823 to 0.90177A
2.3953 to 2.4047A
1. Measure range coupled to source range when simultaneously sourcing and measuring current.
2. As measured by precision digital multimeter. Use closest possible value, and modify reading limits
accordingly if necessary.
Table 15-9
Model 2611/2612 current measurement accuracy limits
Model 2611/2612
source and
measure range
1
Source
current
2
Model 2611/2612
current reading limits
(1 year, 18°C–28°C)
100nA
1µA
10µA
100µA
1mA
10mA
100mA
1A
1.5A
10A
90.000nA
0.9000µA
9.0000µA
90.000µA
0.9000mA
9.0000mA
90.000mA
0.90000A
1.3500A
9.0000A
89.855 to 90.145nA
0.899275 to 0.900725µA
8.99625 to 9.00375µA
89.957 to 90.043µA
0.89962 to 0.90038mA
8.9957 to 9.0043mA
89.962 to 90.038mA
0.89823 to 0.90177A
1.345825 to 1.354175A
8.966 to 9.034A
1. Measure range coupled to source range when simultaneously sourcing and measuring current.
2. As measured by precision digital multimeter. Use closest possible value, and modify reading limits
accordingly if necessary.
15-16 Performance Verification
Series 2600 System SourceMeters Reference Manual