D-10 Status Model
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
Programming and reading registers
Programming enable and transition registers
The only registers that can be programmed by the user are the enable and
transition registers. All other registers in the status structure are read-only
registers. The following explains how to ascertain the parameter values for the
various commands used to program enable registers. The actual commands are
summarized in
.
A command to program an event enable or transition register is sent with a
parameter value that determines the desired state (0 or 1) of each bit in the
appropriate register. The bit positions of the register (
) indicate the
binary parameter value and decimal equivalent. To program one of the registers,
send the decimal value for the bit(s) to be set. The registers are discussed further
in “
Enable and transition registers
” in this appendix, while specific command
parameters to program these registers are outlined later in this section.
Figure D-6
16-bit status register
(2
7
)
(2
6
)
(2
5
)
(2
4
)
(2
3
)
(2
2
)
(2
1
)
(2
0
)
(2
15
) (2
14
) (2
13
) (2
12
) (2
11
) (2
10
)
(2
9
)
(2
8
)
32768 16384
Bit Position
Binary Value
Decimal
Weights
B7
B6
B5
B4
B3
B2
B1
B0
0/1
0/10/1
0/10/1
0/1
0/1
0/1
128
64
32
16
8
4
2
1
B15
B14
B13 B12
B11
B10
B9
B8
8192 4096 2048 1024 512
256
0/1
0/10/1
0/10/1
0/1
0/1
0/1
Bit Position
Binary Value
Decimal
Weights
B. Bits 8 through 15
A. Bits 0 through 7
When using a numeric parameter, registers are programmed by including the
appropriate
<mask>
value, for example:
*ese <mask>
status.standard.enable = <mask>