Series 2600 System SourceMeters Reference Manual
Instrument Control Library 12-117
Return to
2600S-901-01 Rev. A / May 2006
Remarks
• This attribute is used to read or write to the questionable event registers.
• Reading a status register returns a value. The binary equivalent of the returned value
indicates which register bits are set. The least significant bit of the binary number is
bit 0, and the most significant bit is bit 15.
For example, assume value 4352 is returned for the enable register. The binary
equivalent is 0001000100000000. This value indicates that bit B8 (CAL) and bit B12
(OTEMP) are set.
• The used bits of the questionable event registers are described as follows:
•
Bit B8, CAL
– Set bit indicates that calibration is questionable.
•
Bit B12, OTEMP –
Set bit indicates that an over temperature condition was
detected.
•
Bit B13, INST –
Set bit indicates that a bit in the questionable instrument
summary register is set.
Details
See “
” in
.
Example
Sets the OTEMP bit of the questionable enable register:
status.questionable.enable = status.questionable.OTEMP