Series 2600 System SourceMeters Reference Manual
DUT Test Connections 3-13
Return to
2600S-901-01 Rev. A / May 2006
Figure 3-13
HI
LO
DUT
HI
LO
Metal Safety
Shield
SMU
2611/2612
Safety
Earth
Ground
200V
Safety shield for Model 2611/2612 hazardous voltage (200V maximum)
Figure 3-14
CHANNEL A
LO
G
LO
HI G G G HI
S
S
Model 2611/2612
DUT
HI
LO
Safety
Earth
Ground
Chassis
Chassis
Screw
Connections for test circuit shown in
Using shielding and guarding together
shows connections for a test system that uses a noise shield, a safety
shield, and guarding. The guard shields are connected to the driven guard (G) of
the SMU. The noise shield is connected to SMU LO. The safety shield is
connected to the chassis and to a safety earth ground.