Series 2600 System SourceMeters Reference Manual
Source-Measure Concepts
8-9
Return to
2600S-901-01 Rev. A / May 2006
Examples
Example 1:
Using a Model 2611 to charge a 5V battery with 1.5A, while operating at 50
°
C
ambient temperature; what is the maximum duty cycle?
Assuming the 20V range will be used to measure the voltage:
DC
MAX
56 20
–
(
)
38
5
( )
–
(
)
0
( )
–
38
5
( )
–
(
)
1.5
(
)
-------------------------------------------------------------------
100
×
2
≤
DC
MAX
≤ 52.9%
Example 2:
Using a Model 2602 to pulse 10A of current from a bias level of 500mA, into a
very low impedance (100 m
Ω
), while operating at 40
°
C ambient temperature;
what is the maximum duty cycle?
Assuming the 1V range will be used to measure the voltage:
DC
MAX
56 10
–
(
)
18
0.1
(
)
0.5
(
)
–
(
)
0.5
(
)
–
18
10
(
)
0.1
(
)
–
(
)
10
(
)
-------------------------------------------------------------------------------------------
100
×
2
≤
DC
MAX
≤ 4.7%
Example 3:
Using a Model 2612 to charge a 12V battery by sourcing 100mA and then
discharging the battery by sinking 5A, while operating at 35
°
C ambient
temperature; what is the maximum duty cycle?
Assuming the 20V range will be used to measure the voltage:
DC
MAX
56 5
–
(
)
38
12
(
)
–
(
)
0.1
(
)
–
38
12
–
(
)
–
(
)
5
–
(
)
------------------------------------------------------------------------
100
×
2
≤
DC
MAX
≤ 3.7%