13-18 Factory Scripts
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
Waveform:
Pulse initiated here
bias
level
ton
toff
pulse
level
Measurement
(1/NPLC Seconds)
Last source value before pulse train was initiated
Return
Values:
f:
A Boolean flag. This flag will be
true
when the pulse was
successfully configured,
false
when errors were encountered.
msg:
A string message. If the
f
flag is
false
,
msg
will contain an error
message. Otherwise,
msg
will contain a string indicating successful
configuration.
Output
Data:
Pulsed current measurements are stored in the reading buffer specified by
the
buffer
input argument.
Example:
ConfigPulseVMeasureI(smub,
0,
20,
1,
0.001,
0.080,
10,
smub.nvbuffer1, 2)
Setup a pulse train that will use SourceMeter channel B. The pulse amplitude
will be 20V and will return to 0V after 1msec. The pulse will remain at 0V for
80msec and the current limit will be 1A during the pulse. The pulse train will
consist of 10 pulses and the pulse train will be assigned a tag index of 2.
See Also:
Содержание SourceMeter 2600 Series
Страница 507: ...AppendixA Specifications Appendix A topics Model 2600S Specifications page A 2 2600S 901 01 Rev A May 2006...
Страница 575: ...AppendixF DisplayCharacterCodes Appendix F topics Introduction page F 2...
Страница 592: ......
Страница 593: ......