16-14 Calibration
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
4. Be sure to complete steps a through v for all six voltage steps in
or
before continuing with current calibration.
5. Select auto polarity mode:
smua.cal.polarity = smua.CAL_AUTO
Figure 16-2
Connections for current calibration
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
TSP-Link
LINE FUSE
SLOWBLOW
3.15A, 250V
CAT I
S
S
LO LO
HI
HI
G
G
G
G
CHANNEL A
!
!
!
UL
C
US
LISTED
SourceMeter
4ZA4
CAT I
S
S
LO
LO
HI
HI
G G G
G
CHANNEL B
!
R
Digital Multimeter
Amps
Input LO
Model 2602/2612 Channel A
Connections Shown
LO
HI
(100nA to 1A ranges)
Step 3. Current Calibration
1. Connect the Series 2600 to the digital multimeter (see
), and
select the multimeter DC current function.
2. Send this command to initialize current calibration:
smua.source.func = smua.OUTPUT_DCAMPS