12-120 Instrument Control Library
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
status.questionable.instrument.smuX.*
status.questionable.instrument.smuX.condition
status.questionable.instrument.smuX.enable
status.questionable.instrument.smuX.event
status.questionable.instrument.smuX.ntr
status.questionable.instrument.smuX.ptr
* = condition, enable, event, ntr or ptr
smuX = smua or smub
Attribute
Questionable SMU event register sets.
Usage
Reads condition, enable, event, NTR and PTR registers:
quesreg = status.questionable.instrument.smuX.condition
quesreg = status.questionable.instrument.smuX.enable
quesreg = status.questionable.instrument.smuX.event
quesreg = status.questionable.instrument.smuX.ntr
quesreg = status.questionable.instrument.smuX.ptr
Writes to enable, NTR and PTR registers:
status.questionable.instrument.smuX.enable = quesreg
status.questionable.instrument.smuX.ntr = quesreg
status.questionable.instrument.smuX.ptr = quesreg
Set
quesreg
to one of the following values:
0
Clears all bits.
status.questionable.CALIBRATING
Sets CAL bit (B8).
status.questionable.CAL
Sets CAL bit (B8).
status.questionable.OVER_TEMPERATURE
Sets OTEMP bit (B12).
status.questionable.OTEMP
Sets OTEMP bit (B12).
quesreg
can also be set to the decimal weight of the bit to be set. Examples:
To set bit B8 (CAL), set
quesreg
to 256 (2
8
).
To set bit B12 (OTEMP), set
quesreg
to 4096 (2
12
).
To set both bits of the register, set
quesreg
to the sum of the decimal weights. To set
bits B8 and B12, set
quesreg
to 4362
(256 + 4096).
Remarks
• This attribute is used to read or write to the questionable SMU event registers.
• Reading a status register returns a value. The binary equivalent of the returned value
indicates which register bits are set. The least significant bit of the binary number is
bit 0, and the most significant bit is bit 15.
For example, assume value 4362 is returned for the enable register. The binary
equivalent is 0001000100000000. This value indicates that bit B8 (Cal bit) and bit B4
(Meas bit) are set.
• The used bits of the questionable SMU event registers are described as follows:
•
Bit B8, CAL
– Set bit indicates that calibration is questionable.
•
Bit B12, OTEMP –
Set bit indicates that an over temperature condition was
detected.
Details
See “
” in
.
Содержание SourceMeter 2600 Series
Страница 507: ...AppendixA Specifications Appendix A topics Model 2600S Specifications page A 2 2600S 901 01 Rev A May 2006...
Страница 575: ...AppendixF DisplayCharacterCodes Appendix F topics Introduction page F 2...
Страница 592: ......
Страница 593: ......