12-72 Instrument Control Library
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
smuX.measure.filter.count
X = SMU channel (a or b)
Attribute
Number of measured readings to yield one filtered measurement.
Usage
count = smuX.measure.filter.count
-- Reads filter count.
smuX.measure.filter.count = count
-- Writes filter count.
count
Set filter count from 1 to 100.
Remarks
• This attribute is the number of measurements that will be performed to yield one
filtered measurement.
• The
reset
function sets the filter count to 1.
Details
See “
Also see
,
Example
Sets filter count for SMU A:
smua.measure.filter.count = 10
smuX.measure.filter.enable
X = SMU channel (a or b)
Attribute
Enables/disables filtered measurements.
Usage
filter = smuX.measure.filter.enable
-- Reads on/off state of the filter.
smuX.measure.filter.enable = filter
-- Writes on/off state of the filter.
Set
filter
to one of the following values:
0
or
smuX.FILTER_OFF
Disables the filter.
1
or
smuX.FILTER_ON
Enables the filter.
Remarks
• This attribute enables or disables the filter.
• The
reset
function disables the filter.
Details
See “
Also see
Example
Enable the filter for SMU A:
smua.measure.filter.enable = smua.FILTER_ON
smuX.measure.filter.type
X = SMU channel (a or b)
Attribute
Type of filter for measurements.
Usage
type = smuX.measure.filter.type
-- Reads filter type.
smuX.measure.filter.type = type
-- Writes filter type.
Set
type
to one of the following values:
0
or
smuX.FILTER_MOVING_AVG
Selects the moving average filter.
1
or
smuX.FILTER_REPEAT_AVG
Selects the repeat filter.