13-30 Factory Scripts
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
TSP Project Name:
KIFactoryPulse
TSP Test Script Name:
KIPulse
Firmware Version:
1.2.0 and later
Function:
InitiatePulseTest
Usage:
f, msg = InitiatePulseTest(tag)
Description: This function initiates the pulse configuration assigned
tag
.
Parameters: tag:
Numeric identifier of pulse train configuration to be initiated.
Return
Values:
f:
A Boolean flag. This flag will be
true
when the pulse was
successfully configured,
false
when errors were encountered.
msg:
A string message. If the
f
flag is
false
,
msg
will contain an error
message. Otherwise,
msg
will contain a string indicating successful
configuration.
Output
Data:
None
Example(s):
smua.reset()
smua.source.rangev = 5
smua.source.rangei = 1
smua.source.levelv = 0
smua.measure.rangev = 5
smua.measure.rangei = 1
smua.measure.nplc = 0.01
smua.measure.autozero = smua.AUTOZERO_ONCE
smua.nvbuffer1.clear()
smua.nvbuffer1.appendmode = 1
smua.source.output = smua.OUTPUT_ON
f1, msg1 = ConfigPulseVMeasureI(smua, 0, 5, 1, 0.002, 0.2,
10,
smua.nvbuffer1, 1)
if (f1 == true) then
f2, msg2 = InitiatePulseTest(1)
print("Initiate message:", msg2)
else
print("Config errors:", msg1)
end
Содержание SourceMeter 2600 Series
Страница 507: ...AppendixA Specifications Appendix A topics Model 2600S Specifications page A 2 2600S 901 01 Rev A May 2006...
Страница 575: ...AppendixF DisplayCharacterCodes Appendix F topics Introduction page F 2...
Страница 592: ......
Страница 593: ......