8-12 Source-Measure Concepts
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
Figure 8-4
Model 2611/2612 continuous power operating boundaries
-V
+V
-200V
(IV)
Sink
+I
1.5A
100mA
-20V
20V
200V
(III)
Source
(II)
Sink
-100mA
-1.5A
-I
(I)
Source
Energy storage DUT
Energy storage DUT
£
30°C
T
amb
I -Source operating boundaries
Model 2601/2602 I -Source operating boundaries
shows the operating boundaries for the I-Source. Only the first
quadrant of operation is covered. Operation in the other three quadrants is similar.
A shows the output characteristics for the I-Source. As shown, the
Model 2601/2602 SourceMeter can output up to 1.01A at 40V, or 3.03A at 6V.
Note that when sourcing more than 1.01A, voltage is limited to 6V.
B shows the limit lines for the I-Source. The current source limit line
represents the maximum source value possible for the presently selected current
source range. The voltage compliance limit line represents the actual compliance
that is in effect. See “
” earlier in this section. These limit lines are
boundaries that represent the operating limits of the SourceMeter for this quad-
rant of operation. The operating point can be anywhere inside (or on) these limit
lines. The limit line boundaries for the other quadrants are similar.