WARNING
Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30
cm (12 inches) to any part of the scanner, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment
could result. The wireless components in the scanner may be interfered
with by other equipment, even if the other equipment is fully compliant with
CISPR emission requirements.
CAUTION
After using the scanner on a patient, clean the scanner according to the
cleaning instructions found in section "Cleaning / preventive maintenance"
CAUTION
Do not hold the scanner in a way that will cover the cooling vents.
CAUTION
Do not immerse the scanner base in liquid or expose the scanner to
conditions outside the operating conditions, see section "KaVo ProXam iOS
specification" on page 108. Clean the scanner according to the instructions
in section "Cleaning / preventive maintenance" on page 101.
CAUTION
Do not stack the tips near or on any other metal instruments.
CAUTION
Ensure your scanner is properly maintained, see section "Cleaning /
preventive maintenance" on page 101.
CAUTION
If you suspect equipment malfunction or failure, discontinue using the
scanner and contact technical support immediately. Do not attempt to make
any repairs on the scanner.
CAUTION
Read and comply with all safety, warning, and instructional labels on the
scanner.
CAUTION
If the scanner has been operational for an extended period of time, the
surface may become warm to the touch. If the temperature becomes too
warm to the touch within a minute of use, turn it off and allow it to cool. The
surface of the applied part (scanning tip) may reach a temperature of 43°C
during normal operation. This is necessary to prevent fogging of the mirror
and is safe for patient contact.
4.1 Reporting serious incidents
Serious incidents that have occurred in relation to the device must be
reported to the manufacturer and the local competent authority.
4 Safety precautions
Instructions for use
KaVo ProXam iOS 9