Immunity test
IEC 60601 Test level
Compliance level
Electromagnetic
environment -guidance
Voltage dips, short
interruptions and
variations on power
supply input lines
IEC 61000-4-11
0% UT; 0.5 cycle at 0°,
45°, 90°, 135°, 180°,
225°, 270°, and 315°
0% UT; 1 cycle 70%
UT; 25 cycles Single
phase: at 0° 0% UT;
250 cycle
Not applicable
Main power quality
should be that of a
typical commercial or
hospital environment.
If the user of the
BSVD-1000 system
requires continued
operation during power
mains interruptions, it is
recommended that the
scanner be powered
from an uninterruptible
power supply (UPS) or
a battery.
Power frequency
(50/60Hz)
Magnetic field IEC
61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Note:Uт is the A.C. mains voltage prior to application of the test level.
Conducted RF IEC
61000-4-6
3 Vrms 150 kHz to 80
MHz 6 V m) in ISM
bands between 0.15
MHz and 80 MHz
Not applicable
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the scanner,
including cables, than
the recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.