13 Cleaning / preventive maintenance
13.1 Tip infection control procedure
NOTE
For intraoral scanning systems only.
The following instructions are for the removable tip of the scanner, not for the
entire scanner. See "Cleaning scanner base and other surfaces after every
patient" on page 104 for cleaning the base of the scanner.
WARNING
The scanning tips have been manufactured and tested for specific
disinfection protocols. Operational, maintenance and replacement
instructions should be followed for the product to perform as designed.
Cleaning the scanner tip incorrectly may result in incomplete disinfection
and/or permanent damage to the tip.
WARNING
It is of utmost importance to clean the scanner tip after every patient to
ensure proper infection control.
WARNING
Do not place in ultrasonic cleaner.
WARNING
Do not use chemical sterilant.
WARNING
Do not pouch with other instruments.
WARNING
Do not stack the tips near or on any other metal instruments.
WARNING
Autoclave sterilisation cycle times exceeding 10 minutes are excessive and
may damage the tip.
CAUTION
To prevent scratches, avoid brushing the mirror.
CAUTION
Autoclave baskets are not indicated for this cleaning procedure.
13 Cleaning / preventive maintenance
Instructions for use
KaVo ProXam iOS 101