Table 202
Guidance and manufacturer’s declaration – electromagnetic immunity
UNIDRIVE
®
S III model 20 7010 20-1 is intended for use in the electromagnetic environment specified below.
The customer or user of the UNIDRIVE
®
S III model 20 7010 20-1 should ensure that it is used in such an environment.
Immunity test
EN/IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power supply lines
1 kV for input/output lines
2 kV/1 kV
Mains power quality should be that of a typical commercial
or hospital environment.
Surge
IEC 61000-4-5
1 kV line(s) to line(s)
2 kV line(s) to earth
2 kV/1 kV
Mains power quality should be that of a typical commercial
or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycle
70 % UT
(30 % dip in UT)
for 25 cycle
<5 % UT
(>95 % dip in UT)
for 5 s
0 V, 10 msec
92 V, 100 msec
168 V, 500 msec
0 V, 5 sec
Mains power quality should be that of a typical commercial
or hospital environment.
If the user of the equipment or system requires continued
operation during power mains interruptions,
it is recommended that the equipment or system be powe-
red from an uninterruptible power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels cha-
racteristic of a typical location in a typical commercial or
hospital environment.