EasyMatch QC User’s Manual Version 2.1
UltraScan VIS
11G-13
1-inch Port Plate with Glass
This option provides a snap-in port plate with an opening one inch in
diameter that is covered with glass. The glass insert protects the inside of
the instrument from sample and airborne particles.
To install the port insert,
1. Remove the standard port insert by pulling it out of the port insert
retainer. Replace it with the desired port insert, making sure that the
flat portion of the insert is facing outward with the beveled portion toward the interior of the sensor.
2. Standardize the instrument. When using the glass cover, standardize the instrument with the glass in
place. However, you should check measurements of the green and white tiles and perform any other
diagnostics without the glass cover. Difference measurements using the glass cover will be more
accurate than absolute measurements.
4-mm Port Plate
This option provides a modified port plate with a 4-mm round opening.
This port plate should only be used when the small area view option is
available and in use. To install the port plate, remove the standard port
plate and snap the modified port plate into place at the reflectance port.
Standardize the instrument in a reflectance mode using the 0.375-inch
(9.525 mm) area view.
Note: This port plate has an opening that is smaller than the usual 1-inch
port plate. Performance specifications (i.e., factory-supplied color values,
stability, accuracy) are based on the standard 1-inch round port plate.
These specifications will not be met with the modified port plate. The change of geometry that
occurs will cause a slight shift in the color values of the calibration standards.
Reflectance Sample Shelf with Light Cover
The optional reflectance sample shelf with light cover can be placed at the reflectance port for
measurement of powders, granules, pastes, or liquids held in a transmission cell (10-mm, 20-mm, 33-
mm, or 50-mm cell). An opaque cover is also available to prevent ambient room light from affecting the
measurement.
To install the reflectance sample shelf:
1. Lower the sample clamp as far as it will go.
2. Insert the two finger clamps of the sample shelf into the guidance holes located below the reflectance
port on the front of the UltraScan VIS.
Содержание EasyMatch QC
Страница 11: ...Part I Software Overview...
Страница 105: ...EasyMatch QC User s Manual Version 2 1 2 50 Lessons...
Страница 112: ...Part II Software Menu Commands...
Страница 140: ...EasyMatch QC User s Manual Version 2 1 3 28 The File Menu...
Страница 150: ...EasyMatch QC User s Manual Version 2 1 3 38 The File Menu...
Страница 156: ...EasyMatch QC User s Manual Version 2 1 5 2 The View Menu Toolbar is not displayed...
Страница 227: ...Part III Instruments...
Страница 290: ...Part IV Reference...
Страница 325: ...EasyMatch QC User s Manual Version 2 1 Measurement Values A 35...