EasyMatch QC User’s Manual Version 2.1
2-14
Lessons
5.
Remove the black card from the transmission compartment. If you will be measuring a film or other
clear solid sample, simply click
Next
to continue the standardization process. If you will be
measuring a liquid, place a clear liquid (distilled water for water-based samples, toluene or benzene
for resins, or mineral oil for oils) in a cell of the desired size and center the clear portion of the cell
over the transmission port (sphere) in the transmission compartment. It is suggested that you use a
transmission cell holder to do so. Close the transmission compartment door.
6.
Place the white standard tile at the reflectance port (outside the instrument) with the white side
toward the port. (This instruction does not apply to ColorQuest XT.) Then click
Next
to set the top
of scale. Standardization is then complete. Click
Finish
and then
OK
to exit the Modes screen.
Leave the white tile at the reflectance port while you make your measurements.
7.
Center your sample over the transmission port (the hole in the sphere inside the transmission
compartment). If you are measuring a film or flat solid sample, using the transmission clamp to hold
the sample flat against the port is recommended. If you are measuring a liquid sample, fill an
appropriately-sized transmission cell with the liquid and center one of the clear sides of the cell over
the transmission port. Use of a transmission cell holder is recommended. Be sure that the level of
the liquid is sufficient to cover the opening in the sphere and that the cell is flat against the opening.
Close the transmission compartment door.
Note: The size of the transmission cell used for sample measurements must match that of the cell
used for standardization.
8.
Read the sample as a standard or sample by pressing
F2
or
F3
, respectively. Other samples may
then be placed in the transmission compartment and read, if desired.
Содержание EasyMatch QC
Страница 11: ...Part I Software Overview...
Страница 105: ...EasyMatch QC User s Manual Version 2 1 2 50 Lessons...
Страница 112: ...Part II Software Menu Commands...
Страница 140: ...EasyMatch QC User s Manual Version 2 1 3 28 The File Menu...
Страница 150: ...EasyMatch QC User s Manual Version 2 1 3 38 The File Menu...
Страница 156: ...EasyMatch QC User s Manual Version 2 1 5 2 The View Menu Toolbar is not displayed...
Страница 227: ...Part III Instruments...
Страница 290: ...Part IV Reference...
Страница 325: ...EasyMatch QC User s Manual Version 2 1 Measurement Values A 35...