EasyMatch QC User’s Manual Version 2.1
11G-30
UltraScan VIS
6. Complete the standardization by measuring the cell blank as the top of scale.
7. Remove the cell from the cell holder and rinse it with the sample to be measured. Fill the cell with
the sample and measure it the same way you did the blank.
Note: It is important, particularly when using disposable cells, that the same cell used during
standardization be used for sample measurements.