EasyMatch QC User’s Manual Version 2.1
8-18
The Sensor Menu
ColorQuest II Sphere: Pass if dX, dY, and dZ are less than or equal to 0.02 else fail.
ColorQuest 45/0 LAV: Pass if dX, dY, and dZ are less than or equal to 0.02 else fail.
LabScan XE: Pass if dE* is less than or equal to 0.09 else fail.
MiniScan XE Plus: Pass if dX, dY, and dZ are less than or equal to 0.18 else fail.
MiniScan EZ: Pass if dE* is less than or equal to 0.05 else fail.
UltraScan XE: Pass if dL*, da*, and db* are less than or equal to 0.02 else fail.
UltraScan PRO: Pass if dE* is less than or equal to 0.03 else fail.
UltraScan VIS: Pass if dE* is less than or equal to 0.03 else fail.
Click
Yes
to close the Repeatability Job. The following pdf report will appear documenting the results
of the test. As an option, this report can be printed as a hardcopy for your records. This
Repeatability
Test Report
is automatically saved in c:\HunterLab\EasyMatchQC-ER Client\Reports with a unique
data/time stamp for future reference.
File/Exit
to close the report and finish the test
Didymium Filter Test
(available for ColorQuest XE, ColorQuest XT, UltraScan XE, UltraScan PRO,
and UltraScan VIS only)
When
Didymium Filter Test
is selected from the
Diagnostics
submenu, a special job is opened in
EasyMatch QC and the following prompt is shown.
Содержание EasyMatch QC
Страница 11: ...Part I Software Overview...
Страница 105: ...EasyMatch QC User s Manual Version 2 1 2 50 Lessons...
Страница 112: ...Part II Software Menu Commands...
Страница 140: ...EasyMatch QC User s Manual Version 2 1 3 28 The File Menu...
Страница 150: ...EasyMatch QC User s Manual Version 2 1 3 38 The File Menu...
Страница 156: ...EasyMatch QC User s Manual Version 2 1 5 2 The View Menu Toolbar is not displayed...
Страница 227: ...Part III Instruments...
Страница 290: ...Part IV Reference...
Страница 325: ...EasyMatch QC User s Manual Version 2 1 Measurement Values A 35...