EasyMatch QC User’s Manual Version 2.1
UltraScan VIS
11G-29
3. Initiate standardization in TTRAN mode for the 1.000-inch (25.400 mm) area view, sliding the black
card device between the cell holder and the hole in the sphere when prompted for the black card.
4. When prompted for the cell blank, fill the desired analytical cell with distilled water or another clear
liquid you would like to use as your top of scale. Some cells will have a fill line. The liquid should
at least reach this line.
5. Slide the transmission cell into the cell holder as far as it will go. If your cell has an indicator arrow,
point it toward the hole in the instrument sphere.
Fill line
Indicator
arrow