EasyMatch QC User’s Manual Version 2.1
UltraScan VIS
11G-47
Replacing the Optional View Lamp
The view lamp is used to illuminate the sample when using the retroviewer option.
Note: Replace with a specified lamp, HL#A13-1004-022.
To change the view lamp:
1. Disconnect the UltraScan VIS from its power source.
2. Wait two minutes for complete discharge of the capacitors.
3. Disengage and remove the sample clamp.
4. Loosen the two screws on the front cover with a flat screwdriver.
Note: These screws are captive within the instrument cover housing and cannot be removed.
5. Loosen the thumbscrew on the inner panel bezel.