EasyMatch QC User’s Manual Version 2.1
UltraScan VIS
11G-19
5. Close the transmission compartment door and make the measurement.
For measurements in Total Transmission:
1. Position the transmission clamp so that the curved opening is aligned at the opening of the
sphere
at
the front of the transmission compartment. The inner curvature of the clamp should align with the
outer curvature of the sphere.
2. When properly aligned, the holes in the bottom of the clamp assembly will align with holes in the
floor of the transmission compartment. Place a washer and one 1/4-20 7/8-inch screw in each of the
holes and tighten by turning the screw clockwise. (You may finger-tighten the screws or you may
use the Allen wrench that is included with the transmission clamp.)
3. Follow the software prompts to standardize in total transmission (TTRAN).
4. Place the film sample at the sphere opening by pulling back on the tab at the bottom of the clamp
and sliding the transmission clamp holder backward. Place the film at the sphere opening and
release the tab. The rubber grommet at the front of the transmission clamp will firmly hold the
sample in place for measurement.
5. Close the transmission compartment door and make the measurement.
Transmission clamp as
placed in the transmission
compartment for
measurements in regular
transmission. The clamp is
placed flat at the lens.
Transmission clamp as placed
in the transmission
compartment for measurements
in total transmission. The
clamp is placed flat at the
sphere.
Содержание EasyMatch QC
Страница 11: ...Part I Software Overview...
Страница 105: ...EasyMatch QC User s Manual Version 2 1 2 50 Lessons...
Страница 112: ...Part II Software Menu Commands...
Страница 140: ...EasyMatch QC User s Manual Version 2 1 3 28 The File Menu...
Страница 150: ...EasyMatch QC User s Manual Version 2 1 3 38 The File Menu...
Страница 156: ...EasyMatch QC User s Manual Version 2 1 5 2 The View Menu Toolbar is not displayed...
Страница 227: ...Part III Instruments...
Страница 290: ...Part IV Reference...
Страница 325: ...EasyMatch QC User s Manual Version 2 1 Measurement Values A 35...