ET User’s Guide
3-17
Making Measurements
Measuring Transmission Response
NOTE
By convention, cal kit labels in the analyzer indicate the sex of the port
with which they are used. For example, the default cal kit for the
analyzer is type-N female because the front panel RF ports are female
(the calibration standards, in turn, have male connectors).
Calibrating for Insertable Devices
When measuring a through-standard during calibration, the test ports
normally mate directly together. For example, two cables with the
appropriate connectors can be connected directly without a
through-adapter, resulting in a zero-length through-standard. An
insertable device is a device that can be inserted in place of a zero-length
through-standard. It has the same type of connector on each port, but
with a different sex, or the same type of sexless connector on each port
(APC-7, for example).
provides detail about when this calibration is necessary, and
information about other calibrations available for transmission
measurements. If you wish to perform an enhanced response calibration
on your insertable device, perform the following steps:
NOTE
If you are going to be using calibration standards other than the default
(female type-N), you must select the connector type by pressing
and then selecting the appropriate type.
1. Press
.
2. The instrument prompts you to connect four standards—open, short,
load, and through cable—as shown in the following graphic.
CAL
More Cal Cal Kit
CAL
Enhanced Response
Содержание HP 8712ET
Страница 8: ...viii ET User s Guide ...
Страница 17: ...ET User s Guide 1 1 1 Installing the Analyzer ...
Страница 34: ...1 18 ET User s Guide Installing the Analyzer Preventive Maintenance ...
Страница 35: ...ET User s Guide 2 1 2 Getting Started ...
Страница 52: ...2 18 ET User s Guide Getting Started Performing the Operator s Check ...
Страница 53: ...ET User s Guide 3 1 3 Making Measurements ...
Страница 58: ...3 6 ET User s Guide Making Measurements Measuring Devices with Your Network Analyzer Figure 3 3 Block Diagram ...
Страница 118: ...3 66 ET User s Guide Making Measurements Measuring Impedance Magnitude ...
Страница 119: ...ET User s Guide 4 1 4 Using Instrument Functions ...
Страница 129: ...ET User s Guide 4 11 Using Instrument Functions Using Markers Figure 4 6 6 dB Bandwidth Marker Search ...
Страница 131: ...ET User s Guide 4 13 Using Instrument Functions Using Markers Figure 4 7 6 dB Notch Marker Search ...
Страница 134: ...4 16 ET User s Guide Using Instrument Functions Using Markers Figure 4 10 Multi Peak Search Mode ...
Страница 135: ...ET User s Guide 4 17 Using Instrument Functions Using Markers Figure 4 11 Multi Notch Search Mode ...
Страница 137: ...ET User s Guide 4 19 Using Instrument Functions Using Markers Figure 4 12 Marker Statistics Function ...
Страница 139: ...ET User s Guide 4 21 Using Instrument Functions Using Markers Figure 4 13 Marker Flatness Function ...
Страница 141: ...ET User s Guide 4 23 Using Instrument Functions Using Markers Figure 4 14 RF Filter Statistics Function ...
Страница 143: ...ET User s Guide 4 25 Using Instrument Functions Using Markers Figure 4 15 Delta Marker Mode ...
Страница 149: ...ET User s Guide 4 31 Using Instrument Functions Using Limit Testing Figure 4 16 Limit Lines ...
Страница 167: ...ET User s Guide 4 49 Using Instrument Functions Customizing the Display Figure 4 22 The Display Annotation ...
Страница 171: ...ET User s Guide 4 53 Using Instrument Functions Customizing the Display Figure 4 23 Normal Display ...
Страница 172: ...4 54 ET User s Guide Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display ...
Страница 207: ...ET User s Guide 5 1 5 Optimizing Measurements ...
Страница 226: ...5 20 ET User s Guide Optimizing Measurements Measuring Devices with Long Electrical Delay ...
Страница 227: ...ET User s Guide 6 1 6 Calibrating for Increased Measurement Accuracy ...
Страница 235: ...ET User s Guide 6 9 Calibrating for Increased Measurement Accuracy Calibration Choices Figure 6 4 Calibration Choices ...
Страница 266: ...6 40 ET User s Guide Calibrating for Increased Measurement Accuracy Checking the Calibration ...
Страница 267: ...ET User s Guide 7 1 7 Front Rear Panel ...
Страница 269: ...ET User s Guide 7 3 Front Rear Panel Connectors Connectors Figure 7 1 Analyzer Connectors Front Panel ...
Страница 289: ...ET User s Guide 7 23 Front Rear Panel Line Module Figure 7 12 Power Cable and Line Mains Plug Part Numbers ...
Страница 292: ...7 26 ET User s Guide Front Rear Panel Line Module ...
Страница 293: ...ET User s Guide 8 1 8 Hardkey Softkey Reference ...
Страница 395: ...ET User s Guide 9 1 9 Specifications ...
Страница 397: ...ET User s Guide 9 3 Specifications System Performance ...
Страница 443: ...ET User s Guide 10 1 10 Safety and Regulatory Information ...
Страница 449: ...ET User s Guide 10 7 Safety and Regulatory Information Regulatory Information ...
Страница 450: ...10 8 ET User s Guide Safety and Regulatory Information Regulatory Information ...
Страница 451: ...ET User s Guide 11 1 11 Factory Preset State and Memory Allocation ...