ET User’s Guide
Making Measurements
Measuring Devices with Your Network Analyzer
When to Use Amplification
• For accurate measurements, amplification may be needed on the
analyzer's RF OUT port. Use amplification when your test device
requires input power that exceeds the analyzer's maximum specified
output power.
NOTE
If you use an amplifier between the analyzer’s output port and your DUT,
you won’t be able to measure the reflection response of your DUT.
The maximum specified output power is dependent upon the option
configuration of your analyzer as well as the frequency range of your
test setup. It ranges from +6 to +16 dBm. See
to determine the maximum specified output power of
your analyzer.
When to Change the System Impedance
Your analyzer has a system characteristic impedance of either 50 or 75
ohms, yet may be changed to the alternate impedance. If using
minimum-loss pads for impedance conversions, the alternate impedance
should be selected so that the measurement results are displayed
relative to the conversion impedance.
For example, if you have a 50 ohm instrument and are making 75 ohm
measurements, you may be using 50 to 75 ohm minimum-loss pads.
Measurement results can be reported relative to 75 ohms, not 50 ohms, if
the alternate system impedance is selected. This includes marker
readouts, Smith chart results, or SRL impedance computations
(Option 100).
To change the system impedance, press the following keys on the
analyzer:
ο
r
The built-in cal kit selections will be converted to the selected system
impedance.
CAL
More Cal
System Z0
50
Ω
75
Ω
Содержание HP 8712ET
Страница 8: ...viii ET User s Guide ...
Страница 17: ...ET User s Guide 1 1 1 Installing the Analyzer ...
Страница 34: ...1 18 ET User s Guide Installing the Analyzer Preventive Maintenance ...
Страница 35: ...ET User s Guide 2 1 2 Getting Started ...
Страница 52: ...2 18 ET User s Guide Getting Started Performing the Operator s Check ...
Страница 53: ...ET User s Guide 3 1 3 Making Measurements ...
Страница 58: ...3 6 ET User s Guide Making Measurements Measuring Devices with Your Network Analyzer Figure 3 3 Block Diagram ...
Страница 118: ...3 66 ET User s Guide Making Measurements Measuring Impedance Magnitude ...
Страница 119: ...ET User s Guide 4 1 4 Using Instrument Functions ...
Страница 129: ...ET User s Guide 4 11 Using Instrument Functions Using Markers Figure 4 6 6 dB Bandwidth Marker Search ...
Страница 131: ...ET User s Guide 4 13 Using Instrument Functions Using Markers Figure 4 7 6 dB Notch Marker Search ...
Страница 134: ...4 16 ET User s Guide Using Instrument Functions Using Markers Figure 4 10 Multi Peak Search Mode ...
Страница 135: ...ET User s Guide 4 17 Using Instrument Functions Using Markers Figure 4 11 Multi Notch Search Mode ...
Страница 137: ...ET User s Guide 4 19 Using Instrument Functions Using Markers Figure 4 12 Marker Statistics Function ...
Страница 139: ...ET User s Guide 4 21 Using Instrument Functions Using Markers Figure 4 13 Marker Flatness Function ...
Страница 141: ...ET User s Guide 4 23 Using Instrument Functions Using Markers Figure 4 14 RF Filter Statistics Function ...
Страница 143: ...ET User s Guide 4 25 Using Instrument Functions Using Markers Figure 4 15 Delta Marker Mode ...
Страница 149: ...ET User s Guide 4 31 Using Instrument Functions Using Limit Testing Figure 4 16 Limit Lines ...
Страница 167: ...ET User s Guide 4 49 Using Instrument Functions Customizing the Display Figure 4 22 The Display Annotation ...
Страница 171: ...ET User s Guide 4 53 Using Instrument Functions Customizing the Display Figure 4 23 Normal Display ...
Страница 172: ...4 54 ET User s Guide Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display ...
Страница 207: ...ET User s Guide 5 1 5 Optimizing Measurements ...
Страница 226: ...5 20 ET User s Guide Optimizing Measurements Measuring Devices with Long Electrical Delay ...
Страница 227: ...ET User s Guide 6 1 6 Calibrating for Increased Measurement Accuracy ...
Страница 235: ...ET User s Guide 6 9 Calibrating for Increased Measurement Accuracy Calibration Choices Figure 6 4 Calibration Choices ...
Страница 266: ...6 40 ET User s Guide Calibrating for Increased Measurement Accuracy Checking the Calibration ...
Страница 267: ...ET User s Guide 7 1 7 Front Rear Panel ...
Страница 269: ...ET User s Guide 7 3 Front Rear Panel Connectors Connectors Figure 7 1 Analyzer Connectors Front Panel ...
Страница 289: ...ET User s Guide 7 23 Front Rear Panel Line Module Figure 7 12 Power Cable and Line Mains Plug Part Numbers ...
Страница 292: ...7 26 ET User s Guide Front Rear Panel Line Module ...
Страница 293: ...ET User s Guide 8 1 8 Hardkey Softkey Reference ...
Страница 395: ...ET User s Guide 9 1 9 Specifications ...
Страница 397: ...ET User s Guide 9 3 Specifications System Performance ...
Страница 443: ...ET User s Guide 10 1 10 Safety and Regulatory Information ...
Страница 449: ...ET User s Guide 10 7 Safety and Regulatory Information Regulatory Information ...
Страница 450: ...10 8 ET User s Guide Safety and Regulatory Information Regulatory Information ...
Страница 451: ...ET User s Guide 11 1 11 Factory Preset State and Memory Allocation ...