Making Measurements
Making a Power Measurement using Broadband Detection
Press the following keys on the analyzer:
N O T E
This example measurement uses the default instrument parameters for a power measurement. If your
particular power measurement requires specific parameters (such as frequency range, source power
level, number of data points, and sweep
enter them now
3-31
C A U T I O N
Damage to your analyzer will occur if the receiver input power exceeds
or 25 Vdc. The analyzer’s source cannot
exceed this
level, however if your DUT has gain, then attenuation on the RF port
may be necessary. See “When to Use Attenuation and
in a
Measurement Setup” earlier in this chapter for more information.
Содержание 8712C
Страница 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Страница 17: ...I 1 Installing the Analyzer ...
Страница 35: ...I 2 Getting Started ...
Страница 54: ...I 3 Making Measurements ...
Страница 81: ......
Страница 119: ...4 Using Instrument Functions ...
Страница 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Страница 223: ...5 Optimizing Measurements ...
Страница 244: ...6 Calibrating for Increased Measurement Accuracy ...
Страница 280: ...7 Automating Measurements ...
Страница 298: ......
Страница 362: ...8 Front Rear Panel ...
Страница 373: ......
Страница 381: ......
Страница 386: ... 9 w I Suftkey Reference ...
Страница 477: ...I 10 Specifications and Characteristics ...
Страница 501: ...11 Safety and Regulatory Information ...
Страница 508: ...I 12 preset State and Memory Allocation ...
Страница 527: ...Index ...
Страница 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...