I
I
Optimizing Measurements
Increasing Sweep Speed
If averaging has been turned on, (it is off in the preset condition) reducing
the averaging factor (or turning it off altogether) will increase the analyzer’s
measurement speed. Averaging requires multiple sweeps which increases
measurement time. Turning off averaging and using a narrower system
bandwidth may produce faster results.
Press
and enter an averaging factor that is less than
the value displayed on the analyzer screen and press
If you want to turn the averaging off, press
.
‘lb reduce the number of measurement points, press
of Points and use the front panel knob, the
keys, or the
numeric keypad to enter the reduced number.
Generally, as the number of points is decreased, so is the sweep time.
However, other factors will affect the sweep time such as:
l
using frequency bands that contain very low frequencies (below
approximately 20 MHz)
l
the number of band crossing points encountered in a sweep (at
approximately 1900 MHz, 2310 MHz, and 2620 MHz) (HP
only)
l
if the
rate is above the maximum rate at which the source can
be swept
The following graph shows an example of the relationship between the
number of points, frequency span, and sweep time. This graph was created
with data from a setup on an HP
using a center frequency of
1500 MHz, and a system bandwidth setting of medium.
5-5
Содержание 8712C
Страница 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Страница 17: ...I 1 Installing the Analyzer ...
Страница 35: ...I 2 Getting Started ...
Страница 54: ...I 3 Making Measurements ...
Страница 81: ......
Страница 119: ...4 Using Instrument Functions ...
Страница 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Страница 223: ...5 Optimizing Measurements ...
Страница 244: ...6 Calibrating for Increased Measurement Accuracy ...
Страница 280: ...7 Automating Measurements ...
Страница 298: ......
Страница 362: ...8 Front Rear Panel ...
Страница 373: ......
Страница 381: ......
Страница 386: ... 9 w I Suftkey Reference ...
Страница 477: ...I 10 Specifications and Characteristics ...
Страница 501: ...11 Safety and Regulatory Information ...
Страница 508: ...I 12 preset State and Memory Allocation ...
Страница 527: ...Index ...
Страница 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...