I
Calibrating for Increased Measurement
Accuracy
Choose an Appropriate Calibration Method
When using a calibration kit other than one that is compatible with the
standard type-N female connectors
kit model number HP
for
50 analyzers or HP
for 75 analyzers), you can either select
a connector type that is stored the analyzer or input your own
kit
definitions.
The following table shows the connector types that are stored in the analyzer
along with the appropriate cal kit model number.
Connector Type
T y p e - N
Type-N
175
Type-F
3.5
mm
To select the connector type by pressing
and then the
connector type applicable to your measurement.
If you are calibrating with type-F connectors, you must follow the procedure
the
HP
Calibration Kit User
For other connector types, the easy way to enter definitions is to download
them from a file on disk. Use the template program on the Example Programs
disk (DOS format) as a starting point. The format of a
kit
is explained
in the next section.
6-17
Содержание 8712C
Страница 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Страница 17: ...I 1 Installing the Analyzer ...
Страница 35: ...I 2 Getting Started ...
Страница 54: ...I 3 Making Measurements ...
Страница 81: ......
Страница 119: ...4 Using Instrument Functions ...
Страница 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Страница 223: ...5 Optimizing Measurements ...
Страница 244: ...6 Calibrating for Increased Measurement Accuracy ...
Страница 280: ...7 Automating Measurements ...
Страница 298: ......
Страница 362: ...8 Front Rear Panel ...
Страница 373: ......
Страница 381: ......
Страница 386: ... 9 w I Suftkey Reference ...
Страница 477: ...I 10 Specifications and Characteristics ...
Страница 501: ...11 Safety and Regulatory Information ...
Страница 508: ...I 12 preset State and Memory Allocation ...
Страница 527: ...Index ...
Страница 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...