Making Measurements
Measuring Reflection Response
Your analyzer can provide highly accurate measurements without performing
any additional user-calibrations if certain conditions are met. This example
describes how to perform a reflection one-port calibration. A one port
calibration uses known standards to correct for directivity, source match, and
frequency response errors in narrowband measurements.
To perform a reflection one-port calibration you will need one of the following
calibration kits depending on the nominal impedance of your analyzer:
HP 850323
for 50 type-N female connector calibrations
HP
for type-N female or type-N male 50 connector calibrations
HP 850363
for 75 type-N female connector calibrations
HP
for type-N female or type-N male 75 connector calibrations
HP
for 3.5 mm female or 3.5 mm male 50 connector
calibrations
HP
for Type-F 75 connector calibrations
N O T E
If you are going to be using calibration standards other than the default (female
you must
select the type of connector type by pressing
and then selecting the appropriate
3-25
Содержание 8712C
Страница 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Страница 17: ...I 1 Installing the Analyzer ...
Страница 35: ...I 2 Getting Started ...
Страница 54: ...I 3 Making Measurements ...
Страница 81: ......
Страница 119: ...4 Using Instrument Functions ...
Страница 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Страница 223: ...5 Optimizing Measurements ...
Страница 244: ...6 Calibrating for Increased Measurement Accuracy ...
Страница 280: ...7 Automating Measurements ...
Страница 298: ......
Страница 362: ...8 Front Rear Panel ...
Страница 373: ......
Страница 381: ......
Страница 386: ... 9 w I Suftkey Reference ...
Страница 477: ...I 10 Specifications and Characteristics ...
Страница 501: ...11 Safety and Regulatory Information ...
Страница 508: ...I 12 preset State and Memory Allocation ...
Страница 527: ...Index ...
Страница 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...