Automating Measurements
Displaying Measurement Results
The active marker’s value is displayed in the upper right area of the graticule.
If marker bandwidth (or notch) search is selected, the bandwidth (or notch)
information is displayed instead. This marker information can be used to
view exact measured data at critical frequency points. Note the customized
measurement channel notation in Figure 11. In addition to the active
marker’s readout, four of the marker’s values are displayed in the
area during front panel use. This makes it easy to quickly read the measured
data at several marker positions.
The triangular marker symbols can also be used to graphically indicate critical
frequency points of the measurement. For example, a marker can be set at
the desired center frequency for a notch filter, and the operator can tune the
filter until the notch is at the same frequency as the marker.
Marker search types include:
max search
l
mm search
l
target search
l
bandwidth
l
notch
multi peak
multi notch
When marker tracking is turned on, these searches will be automatically
performed at the end of each sweep. This can be useful in tuning
applications.
Other marker functions that can be useful are the marker math functions:
statistics
flatness
l
RF filter stats
7 - 6 9
Содержание 8712C
Страница 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Страница 17: ...I 1 Installing the Analyzer ...
Страница 35: ...I 2 Getting Started ...
Страница 54: ...I 3 Making Measurements ...
Страница 81: ......
Страница 119: ...4 Using Instrument Functions ...
Страница 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Страница 223: ...5 Optimizing Measurements ...
Страница 244: ...6 Calibrating for Increased Measurement Accuracy ...
Страница 280: ...7 Automating Measurements ...
Страница 298: ......
Страница 362: ...8 Front Rear Panel ...
Страница 373: ......
Страница 381: ......
Страница 386: ... 9 w I Suftkey Reference ...
Страница 477: ...I 10 Specifications and Characteristics ...
Страница 501: ...11 Safety and Regulatory Information ...
Страница 508: ...I 12 preset State and Memory Allocation ...
Страница 527: ...Index ...
Страница 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...