Automating Measurements
Controlling Peripherals
N O T E
When using the
15
or
command, the Printer-Select Line is
set High. However, when the instrument is doing hardcopy, the Printer-Select Line is set low The
Printer-Select line may or may not be used by individual printers. Check with your printer manual.
write the value 52 decimal (34 hex, 0011 0100 binary) to the parallel
port’s output pins, use one of the following commands:
OUTPUT
When the write command is executed, the parallel port’s data lines (pins 2-9)
will be set to the specified value, and then a pulse of at least 1 duration
will occur on the strobe line (pin 1). A data setup and data hold time of at
least 1 are guaranteed. See Figure 7-9.
7-56
Содержание 8712C
Страница 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Страница 17: ...I 1 Installing the Analyzer ...
Страница 35: ...I 2 Getting Started ...
Страница 54: ...I 3 Making Measurements ...
Страница 81: ......
Страница 119: ...4 Using Instrument Functions ...
Страница 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Страница 223: ...5 Optimizing Measurements ...
Страница 244: ...6 Calibrating for Increased Measurement Accuracy ...
Страница 280: ...7 Automating Measurements ...
Страница 298: ......
Страница 362: ...8 Front Rear Panel ...
Страница 373: ......
Страница 381: ......
Страница 386: ... 9 w I Suftkey Reference ...
Страница 477: ...I 10 Specifications and Characteristics ...
Страница 501: ...11 Safety and Regulatory Information ...
Страница 508: ...I 12 preset State and Memory Allocation ...
Страница 527: ...Index ...
Страница 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...