I
Making Measurements
Measuring Group Delay
1.
2.
3.
Since we are measuring the transmission group delay, a transmission
calibration can be performed to improve accuracy
The instrument prompts you to connect four standards-open, short, load,
and through cable-as shown below.
NETWORK ANALYZER
N E T W O R K A N A L Y Z E R
R F O U T
R F O U T
R F I N
OPEN
SHORT
LOAD
Open, Short, load Connections
Through Cable Connection
4. Press Measure
after connecting each standard.
5. The analyzer will measure each standard and then calculate the new
calibration coefficients. The message “Calibration complete. will
appear for a few seconds when the analyzer is done calculating the new
error correction array.
6. The calibration may be saved in memory or on a disk for later use if you
wish. See Chapter 6 for information on saving calibrations.
3-51
Содержание 8712C
Страница 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Страница 17: ...I 1 Installing the Analyzer ...
Страница 35: ...I 2 Getting Started ...
Страница 54: ...I 3 Making Measurements ...
Страница 81: ......
Страница 119: ...4 Using Instrument Functions ...
Страница 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Страница 223: ...5 Optimizing Measurements ...
Страница 244: ...6 Calibrating for Increased Measurement Accuracy ...
Страница 280: ...7 Automating Measurements ...
Страница 298: ......
Страница 362: ...8 Front Rear Panel ...
Страница 373: ......
Страница 381: ......
Страница 386: ... 9 w I Suftkey Reference ...
Страница 477: ...I 10 Specifications and Characteristics ...
Страница 501: ...11 Safety and Regulatory Information ...
Страница 508: ...I 12 preset State and Memory Allocation ...
Страница 527: ...Index ...
Страница 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...