Making Measurements
Measuring Transmission Response
4. The analyzer will measure each standard and then calculate the new
calibration coefficients. The message “Calibration complete. will
appear for a few seconds when the analyzer is done calculating the new
error correction array.
5. The calibration may be saved in memory or on a disk for later use if you
wish. However, the current calibration for each measurement channel is
always saved in nonvolatile (battery-backed) memory and will be used
the next time the analyzer is preset (see note below) or turned on. See
Chapter 6 for information on saving calibrations to the analyzer’s internal
memory, the analyzer’s random access memory (RAM), or to a floppy disk.
N O T E
Changing sweep frequencies land other source parameters) may affect your calibration. See Chapter 6,
“Calibrating for Increased Measurement Accuracy,” for more information.
3-20
I
I
Содержание 8712C
Страница 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Страница 17: ...I 1 Installing the Analyzer ...
Страница 35: ...I 2 Getting Started ...
Страница 54: ...I 3 Making Measurements ...
Страница 81: ......
Страница 119: ...4 Using Instrument Functions ...
Страница 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Страница 223: ...5 Optimizing Measurements ...
Страница 244: ...6 Calibrating for Increased Measurement Accuracy ...
Страница 280: ...7 Automating Measurements ...
Страница 298: ......
Страница 362: ...8 Front Rear Panel ...
Страница 373: ......
Страница 381: ......
Страница 386: ... 9 w I Suftkey Reference ...
Страница 477: ...I 10 Specifications and Characteristics ...
Страница 501: ...11 Safety and Regulatory Information ...
Страница 508: ...I 12 preset State and Memory Allocation ...
Страница 527: ...Index ...
Страница 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...