Using Instrument Functions
Using limit Testing
Flatness
1. This limit test requires that you
define a segment on the measurement
trace using markers 1 and 2 (or markers 3 and 4 for measurement channel
2). Then press
to enable the statistics marker search. See
Use Marker Flatness”
earlier in this chapter for more information on using marker statistics.
2. Press
Limit
Limits .
3. Use the front panel knob or the
keys to select Flatness in the
marker limit test table. Turn this limit function on by pressing the
Limit
OFF key. Note that the entry in the on/off column of the
table changes to “on.
4. Press
Limit
and enter the maximum
limit using the front panel keypad and terminating the entry with the
k e y .
Press Edit Limit
Limit and enter the minimum limit
using the front panel keypad and terminating the entry with the
key.
N O T E
Note that there are no visible limit lines or indicators on the display with this limit test function.
Therefore the
Limit
OFF
function has no effect when using marker limits.
4-41
Содержание 8712C
Страница 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Страница 17: ...I 1 Installing the Analyzer ...
Страница 35: ...I 2 Getting Started ...
Страница 54: ...I 3 Making Measurements ...
Страница 81: ......
Страница 119: ...4 Using Instrument Functions ...
Страница 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Страница 223: ...5 Optimizing Measurements ...
Страница 244: ...6 Calibrating for Increased Measurement Accuracy ...
Страница 280: ...7 Automating Measurements ...
Страница 298: ......
Страница 362: ...8 Front Rear Panel ...
Страница 373: ......
Страница 381: ......
Страница 386: ... 9 w I Suftkey Reference ...
Страница 477: ...I 10 Specifications and Characteristics ...
Страница 501: ...11 Safety and Regulatory Information ...
Страница 508: ...I 12 preset State and Memory Allocation ...
Страница 527: ...Index ...
Страница 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...