Automating Measurements
Saving Measurement Results
Similar to Save ASCII, the analyzer can also save the measurement data onto
disk as an instrument state
Use
, and turn Data ON and
turn Inst State and Cal OFF
Tip
A file saved in this manner is smaller than a file saved using
.
Save
Data
uses 6 bytes per point, as opposed to about 20 bytes per point for
. However, the file type is binary, and contains a header, making it difficult to read.
The analyzer can measure a
and compute its center frequency,
bandwidth, Q, and Loss. You can query this information using the SCPI
command
CALC:MARK:FUNC:RES?
For example:
10 OUTPUT
na;
: MARK : BWID
! -3 bandwidth
20 OUTPUT
!
Get
result of bandwidth search
30 ENTER
For more details, refer to the “Example Programs” chapter of the
Programmer’s Guide.
7-74
Содержание 8712C
Страница 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Страница 17: ...I 1 Installing the Analyzer ...
Страница 35: ...I 2 Getting Started ...
Страница 54: ...I 3 Making Measurements ...
Страница 81: ......
Страница 119: ...4 Using Instrument Functions ...
Страница 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Страница 223: ...5 Optimizing Measurements ...
Страница 244: ...6 Calibrating for Increased Measurement Accuracy ...
Страница 280: ...7 Automating Measurements ...
Страница 298: ......
Страница 362: ...8 Front Rear Panel ...
Страница 373: ......
Страница 381: ......
Страница 386: ... 9 w I Suftkey Reference ...
Страница 477: ...I 10 Specifications and Characteristics ...
Страница 501: ...11 Safety and Regulatory Information ...
Страница 508: ...I 12 preset State and Memory Allocation ...
Страница 527: ...Index ...
Страница 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...