Making Measurements
Measuring Transmission Response
Your analyzer can provide highly accurate measurements without performing
any additional user-calibrations if certain conditions are met. This example
describes how to perform an enhanced transmission response calibration.
When you perform an enhanced transmission response calibration, the
analyzer performs correction at the selected number of data points across
the selected frequency band. Interpolation recalculates the error correction
array for reduced frequency spans. If the frequency span is increased the
calibration is invalidated, and the default response calibration is automatically
restored.
Chapter 6 provides detail about when additional calibration is necessary, and
information about other calibrations available for transmission measurements.
If you wish to calibrate your instrument for a transmission response
measurement, perform the following steps:
Press
.
2. The instrument prompts you to connect four standards-open, short, load,
and through cable-as shown below.
NETWORK ANALYZER
N E T W O R K A N A L Y Z E R
R F O U T
R F O U T
RF IN
OPEN
SHORT
LOAD
Open, Short, load Connections
Through
3. Press
after connecting each standard.
3-19
Содержание 8712C
Страница 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Страница 17: ...I 1 Installing the Analyzer ...
Страница 35: ...I 2 Getting Started ...
Страница 54: ...I 3 Making Measurements ...
Страница 81: ......
Страница 119: ...4 Using Instrument Functions ...
Страница 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Страница 223: ...5 Optimizing Measurements ...
Страница 244: ...6 Calibrating for Increased Measurement Accuracy ...
Страница 280: ...7 Automating Measurements ...
Страница 298: ......
Страница 362: ...8 Front Rear Panel ...
Страница 373: ......
Страница 381: ......
Страница 386: ... 9 w I Suftkey Reference ...
Страница 477: ...I 10 Specifications and Characteristics ...
Страница 501: ...11 Safety and Regulatory Information ...
Страница 508: ...I 12 preset State and Memory Allocation ...
Страница 527: ...Index ...
Страница 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...