Calibrating for Increased Measurement
Accuracy
Choose an Appropriate Calibration Method
Normalization is the simplest type of calibration. The analyzer stores data
into memory and divides subsequent measurements by the stored data to
remove frequency response errors.
Follow these general steps when performing a normalization calibration:
1.
2.
3.
4.
Setup the analyzer for your measurement:
Select the type of measurement
Enter operating parameters other than the default
Connect your equipment as you would for an actual measurement, but
omit the DUT.
Notice that the top of the display on the analyzer now shows the type
of measurement followed by
to indicate that the displayed data is
actually the measurement divided by memory.
C A U T I O N
The normalization cal will be invalidated if any frequency settings are
changed after calibration. There is no correction interpolation with a
normalization cal.
C A U T I O N
If you decide to perform any other type of calibration, be sure to turn off
normalization
by pressing
Note that the
indicator
disappears when normalization is turned off.
6-10
Содержание 8712C
Страница 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Страница 17: ...I 1 Installing the Analyzer ...
Страница 35: ...I 2 Getting Started ...
Страница 54: ...I 3 Making Measurements ...
Страница 81: ......
Страница 119: ...4 Using Instrument Functions ...
Страница 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Страница 223: ...5 Optimizing Measurements ...
Страница 244: ...6 Calibrating for Increased Measurement Accuracy ...
Страница 280: ...7 Automating Measurements ...
Страница 298: ......
Страница 362: ...8 Front Rear Panel ...
Страница 373: ......
Страница 381: ......
Страница 386: ... 9 w I Suftkey Reference ...
Страница 477: ...I 10 Specifications and Characteristics ...
Страница 501: ...11 Safety and Regulatory Information ...
Страница 508: ...I 12 preset State and Memory Allocation ...
Страница 527: ...Index ...
Страница 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...