Making Measurements
Measuring Reflection Response
Chapter 6 provides detail about when this calibration is necessary. If you
wish to calibrate your instrument for a reflection one-port measurement,
perform the following steps:
Press (CAL)
2. The instrument will prompt you to connect three standards (open, short
and load) and measure them. See Figure 3-7.
N E T W O R K A N A L Y Z E R
O P E N
S H O R T
L O A D
Figure
Equipment Setup For a Reflection Response Calibration
3. Press
Standard after connecting each standard.
4. The analyzer will measure each standard and then calculate new
calibration coefficients. The message “Calibration complete. will
appear for a few seconds when the analyzer is done calculating the new
error correction array.
5. The calibration may be saved in memory for later use if you wish. See
Chapter 6 for information on saving calibrations.
3-26
Содержание 8712C
Страница 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Страница 17: ...I 1 Installing the Analyzer ...
Страница 35: ...I 2 Getting Started ...
Страница 54: ...I 3 Making Measurements ...
Страница 81: ......
Страница 119: ...4 Using Instrument Functions ...
Страница 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Страница 223: ...5 Optimizing Measurements ...
Страница 244: ...6 Calibrating for Increased Measurement Accuracy ...
Страница 280: ...7 Automating Measurements ...
Страница 298: ......
Страница 362: ...8 Front Rear Panel ...
Страница 373: ......
Страница 381: ......
Страница 386: ... 9 w I Suftkey Reference ...
Страница 477: ...I 10 Specifications and Characteristics ...
Страница 501: ...11 Safety and Regulatory Information ...
Страница 508: ...I 12 preset State and Memory Allocation ...
Страница 527: ...Index ...
Страница 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...