Using Instrument Functions
Using limit Testing
A sloping limit line has different values for its begin and end limits. For
example, create a sloping limit line between 130 MHz and 155 MHz with a
beginning level of -35
and an ending level of -3
Press Prior
Menu
Add
Add
Press End Frequency
.
Press
Press
Limit
N O T E
When creating limit lines, you can enter frequency and limit values using the front panel knob or the
keys as well as the numeric keypad.
Figure 4-16 shows limit lines created that dictate the specified shape of a
In this example, a Elter was connected, and tuned to fall
within the created limit lines. The pass/fail indicator gives constant feedback
regarding status.
4-35
Содержание 8712C
Страница 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Страница 17: ...I 1 Installing the Analyzer ...
Страница 35: ...I 2 Getting Started ...
Страница 54: ...I 3 Making Measurements ...
Страница 81: ......
Страница 119: ...4 Using Instrument Functions ...
Страница 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Страница 223: ...5 Optimizing Measurements ...
Страница 244: ...6 Calibrating for Increased Measurement Accuracy ...
Страница 280: ...7 Automating Measurements ...
Страница 298: ......
Страница 362: ...8 Front Rear Panel ...
Страница 373: ......
Страница 381: ......
Страница 386: ... 9 w I Suftkey Reference ...
Страница 477: ...I 10 Specifications and Characteristics ...
Страница 501: ...11 Safety and Regulatory Information ...
Страница 508: ...I 12 preset State and Memory Allocation ...
Страница 527: ...Index ...
Страница 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...